Development and validation of a capillary high-performance liquid chromatography/electrospray tandem mass spectrometric method for the quantification of bisphenol A in air samples.

Journal: Rapid Communications In Mass Spectrometry : RCM
Published:
Abstract

Bisphenol A (BPA) is a toxic industrial chemical that affects the endocrine system even at low concentrations. A new method, based on capillary high-performance liquid chromatography/electrospray ionization tandem mass spectrometry (HPLC/ESI-MS/MS) analysis, has been developed to determine BPA in atmospheric samples. The method involves collection of air samples (typically 2 m(3)) on glass fiber filters, with ultrasonic extraction and sample concentration under vacuum before analysis. HPLC analysis was performed isocratically at a flow rate of 10 microL min(-1) using a capillary reversed-phase column and MS/MS analysis in negative ion multiple reaction monitoring (MRM) mode, using BPA-d(16) as internal standard. The present method provides linear response in the range 0.007-3.5 microg/filter (R(2) > 0.999) and is characterized by high accuracy (mean bias 2%) and good reproducibility (mean RSD 5%). High sensitivity (LOD = 2 ng/m(3) based on 2 m(3) of air collected), specificity, and speed of the analysis make the present method suitable for routine determination of BPA in the atmosphere, both for ambient and personnel monitoring.

Authors
Laura Sabatini, Anna Barbieri, Francesco Violante