Novel percutaneous catheter thrombectomy in acute massive pulmonary embolism: rotational bidirectional thrombectomy (ROBOT).

Journal: Catheterization And Cardiovascular Interventions : Official Journal Of The Society For Cardiac Angiography & Interventions
Published:
Abstract

Background: Although thrombolysis is a standard therapy in cases of pulmonary embolism (PE), fatal outcome is often observed. We designed and investigated the efficacy of a novel percutaneous catheter therapy, rotational bidirectional thrombectomy (ROBOT), for PE.

Results: Eighteen patients with acute massive PE (Miller score > or = 20) were included in this study. We separated them into two groups [group A (n = 10), thrombolysis; group B (n = 8): thrombolysis and ROBOT or ROBOT alone]. There was no difference in the hemodynamic indices between the groups at diagnosis. ROBOT was designed to fragment emboli by rotating a regular pigtail catheter. Three deaths occurred in group A because of hemodynamic impairment, but there was no death in group B. One day after treatment, systolic pulmonary artery pressure had decreased from 53 +/- 8 to 30 +/- 8 mm Hg (P < 0.05) in group B and from 54 +/- 5 to 42 +/- 19 mm Hg (NS) in group A. The hospitalization period in group B was shorter than that in group A (17 +/- 6 vs. 27 +/- 10 days, P < 0.05).

Conclusions: ROBOT therapy results in a significant, rapid improvement in the hemodynamic situation and in a better outcome than conventional therapy in patients with acute massive pulmonary embolism.

Authors
Masashi Yoshida, Ichiro Inoue, Takuji Kawagoe, Masaharu Ishihara, Yuji Shimatani, Satoshi Kurisu, Kengo Kusano, Tohru Ohe
Relevant Conditions

Thrombectomy, Pulmonary Embolism