Automated liquid-liquid extraction based on 96-well plate format in conjunction with ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) for the quantitation of methoxsalen in human plasma.

Journal: Journal Of Chromatography. B, Analytical Technologies In The Biomedical And Life Sciences
Published:
Abstract

A sensitive, specific and high throughput bioanalytical method using automated sample processing via 96-well plate liquid-liquid extraction and ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) has been developed for the determination of methoxsalen in human plasma. Plasma samples with ketoconazole as internal standard (IS) were prepared by employing 0.2 mL human plasma in ethyl acetate:dichloromethane (80:20, v/v). The chromatographic separation was achieved on a Waters Acquity UPLC BEH C18 column using isocratic mobile phase, consisting of 10 mM ammonium formate and acetonitrile (60:40, v/v), at a flow rate of 0.5 mL/min. The linear dynamic range was established over the concentration range 1.1-213.1 ng/mL for methoxsalen. The method was rugged and rapid with a total run time of 1.5 min. It was successfully applied to a pivotal bioequivalence study in 12 healthy human subjects after oral administration of 10 mg extended release methoxsalen formulation under fasting condition.

Authors
Manish Yadav, Pritesh Contractor, Vivek Upadhyay, Ajay Gupta, Swati Guttikar, Puran Singhal, Sailendra Goswami, Pranav Shrivastav