Objective: This study investigated the reliability, accuracy, and time efficiency of automated hearing assessment using a computer-based telemedicine-compliant audiometer.

Methods: Thirty normal-hearing subjects and eight hearing-impaired subjects were tested with pure-tone air conduction audiometry (125-8,000 Hz) in a manual and automated configuration in a counterbalanced manner. For the normal-hearing group each test was repeated to determine test-retest reliability and recording time, and preference for threshold-seeking method (manual vs. automated) was documented.

Results: Test-retest thresholds were not significantly different for manual and automated testing. Manual audiometry test-retest correspondence was 5 dB or less in 88% of thresholds compared to 91% for automated audiometry. Thresholds for automated audiometry did not differ significantly from manual audiometry with 87% of thresholds in the normal-hearing group and 97% in the hearing-impaired group, corresponding within 5 dB or less of each other. The largest overall average absolute difference across frequencies was 3.6 +/- 3.9 dB for the normal-hearing group and 3.3 +/- 2.4 for the hearing-impaired group. Both techniques were equally time efficient in the normal-hearing population, and 63% of subjects preferred the automated threshold-seeking method.

Conclusions: Automated audiometry provides reliable, accurate, and time-efficient hearing assessments for normal-hearing and hearing-impaired adults. Combined with an asynchronous telehealth model it holds significant potential for reaching underserved areas where hearing health professionals are unavailable.

Authors
De Swanepoel, Shadrack Mngemane, Silindile Molemong, Hilda Mkwanazi, Sizwe Tutshini
Relevant Conditions

Hearing Loss