An analysis of the reason for HLA-C allele dropout in five samples by sequence-based typing

Journal: Zhonghua Yi Xue Yi Chuan Xue Za Zhi = Zhonghua Yixue Yichuanxue Zazhi = Chinese Journal Of Medical Genetics
Published:
Abstract

Objective: To analyze the possible reason for HLA-C allele dropout in routine sequence-based typing (SBT) and improve the accuracy of HLA-C SBT test.

Methods: A total of 620 randomly selected samples from healthy voluntary blood donors in Shenzhen were typed at HLA-C locus by sequence-based typing using the AlleleSEQR HLA-C plus sequence-based typing kit. Samples with no full match result were subjected to cloning and haplotype sequencing of the full-length HLA-C gene. If no novel mutations were found, samples were then retyped, using our self-designed PCR primer pair and PCR conditions replacing the AlleleSEQR HLA-C PCR reagents in the PCR set-up procedure so as to analyze the potential reasons for causing abnormal SBT result.

Results: In the 620 samples typed at HLA-C locus using the AlleleSEQR HLA-C SBT commercial kit, 5 samples with no full match result were identified. The closest genotype showed one nucleotide mismatch with many different allele groups at different nucleotide position. Based on the PCR-SBT nucleotide sequence, heterozygous nucleotides were determined only in exon 4, whereas the nucleotides in exon 2 and 3 were all homozygotes. The results showed that HLA-Cw*0706 allele dropout existed in all the 5 samples with abnormal SBT results initially identified by AlleleSEQR HLA-C SBT kit, no novel mutation was found.

Conclusions: The results indicate that the PCR primer pair incompatible with DNA template may result in allele dropout in HLA-C SBT test. Based on the characterization of HLA-C full-length, it is essential to develop HLA-C SBT kit suitable for Chinese population in the future.

Authors
Jian-qiang Zeng, Yun-ping Xu, Da-ming Wang, Hong-yan Zou, Zhi-hui Deng, Bao-cheng Yang