The ECG lead I paradox in cardiac resynchronization therapy.
Background: In cardiac resynchronization therapy (CRT), the morphology of the QRS complex plays an important role in the determination of the pacing site and effectiveness of stimulation.
Methods: Review of the electrocardiograms (ECGs) of 737 patients with a CRT device showed a negative QRS complex in lead I during right ventricular (RV) pacing and a positive QRS complex during left ventricular (LV) pacing in four patients. The RV lead was positioned in the high RV septum and the coronary sinus leads in a posterior or postero-lateral basal level. Reversed ECG lead or pacemaker lead connection, anodal RV stimulation, and scar tissue-related depolarization abnormalities were excluded as possible causes.
Conclusions: Pacing from the high RV septum may rarely lead to a negative QRS complex and basal positions of the LV lead to a positive QRS complex in lead I during LV pacing. The lead I paradox becomes obvious when both phenomena, that are not interrelated, are present in the same patient.