Automated ankle-brachial pressure index measurement by clinical staff for peripheral arterial disease diagnosis in nondiabetic and diabetic patients.
Objective: Peripheral arterial disease (PAD) is a prognostic marker in cardiovascular disease. The use of Doppler-measured ankle-brachial pressure index (Dop-ABI) for PAD diagnosis is limited because of time, required training, and costs. We assessed automated oscillometric measurement of the ankle-brachial pressure index (Osc-ABI) by nurses and clinical staff.
Methods: Clinical staff obtained Osc-ABI with an automated oscillometric device in 146 patients (83 with diabetes) at the time of Dop-ABI measurement and ultrasound evaluation.
Results: Measurements were obtained in most legs (Dop-ABI 98%; Osc-ABI 95.5%). Dop- and Osc-ABI were significantly related in diabetic and nondiabetic patients with good agreement over a wide range of values. When Dop-ABI
Conclusions: PAD screening could be improved by using Osc-ABI measured by clinical staff with the benefit of greater cost-effectiveness but at the risk of lower diagnostic performance in diabetic patients.