Development and validation of a LC-MS/MS method for the determination of clebopride and its application to a pharmacokinetics study in healthy Chinese volunteers.

Journal: Journal Of Chromatography. B, Analytical Technologies In The Biomedical And Life Sciences
Published:
Abstract

A sensitive and specific liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS/MS) method has been developed and validated for the identification and quantification of clebopride in human plasma using itopride as an internal standard. The method involves a simple liquid-liquid extraction. The analytes were separated by isocratic gradient elution on a CAPCELL MG-III C(18) (5 microm, 150 mm x 2.1 mm i.d.) column and analyzed in multiple reaction monitoring (MRM) mode with positive electrospray ionization (ESI) interface using the respective [M+H](+) ions, m/z 373.9-->m/z184.0 for clebopride, m/z 359.9-->m/z71.5 for itopride. The method was validated over the concentration range of 69.530-4450.0 pg/ml for clebopride. Within- and between-batch precision (RSD%) was all within 6.83% and accuracy ranged from -8.16 to 1.88%. The LLOQ was 69.530 pg/ml. The extraction recovery was on an average 77% for clebopride. The validated method was used to study the pharmacokinetics profile of clebopride in human plasma after oral administration of clebopride.

Authors
Zhirong Tan, Dongsheng Ouyang, Yao Chen, Gan Zhou, Shan Cao, Yicheng Wang, Xiujuan Peng, Honghao Zhou