Optimal degree of protonation for ¹H detection of aliphatic sites in randomly deuterated proteins as a function of the MAS frequency.

Journal: Journal Of Biomolecular NMR
Published:
Abstract

The (1)H dipolar network, which is the major obstacle for applying proton detection in the solid-state, can be reduced by deuteration, employing the RAP (Reduced Adjoining Protonation) labeling scheme, which yields random protonation at non-exchangeable sites. We present here a systematic study on the optimal degree of random sidechain protonation in RAP samples as a function of the MAS (magic angle spinning) frequency. In particular, we compare (1)H sensitivity and linewidth of a microcrystalline protein, the SH3 domain of chicken α-spectrin, for samples, prepared with 5-25 % H(2)O in the E. coli growth medium, in the MAS frequency range of 20-60 kHz. At an external field of 19.96 T (850 MHz), we find that using a proton concentration between 15 and 25 % in the M9 medium yields the best compromise in terms of sensitivity and resolution, with an achievable average (1)H linewidth on the order of 40-50 Hz. Comparing sensitivities at a MAS frequency of 60 versus 20 kHz, a gain in sensitivity by a factor of 4-4.5 is observed in INEPT-based (1)H detected 1D (1)H,(13)C correlation experiments. In total, we find that spectra recorded with a 1.3 mm rotor at 60 kHz have almost the same sensitivity as spectra recorded with a fully packed 3.2 mm rotor at 20 kHz, even though ~20× less material is employed. The improved sensitivity is attributed to (1)H line narrowing due to fast MAS and to the increased efficiency of the 1.3 mm coil.

Authors
Sam Asami, Kathrin Szekely, Paul Schanda, Beat Meier, Bernd Reif