Improved strut coverage and less late incomplete apposition with thin-strut TAXUS Liberté vs. TAXUS Express: the importance of stent platform design for drug-eluting stents.
Background: The objective of this intravascular ultrasound (IVUS) analysis was to evaluate the vascular response of the thin-strut TAXUS Liberté stent compared with the otherwise identical TAXUS Express stent.
Methods: The TAXUS ATLAS and TAXUS ATLAS Long Lesion studies are nonrandomized trials comparing the thin-strut TAXUS Liberté stent to historical TAXUS Express controls from the TAXUS IV and TAXUS V trials. A total of 377 patients enrolled in the two TAXUS ATLAS studies were randomly selected for the IVUS subset and compared to 314 TAXUS Express IVUS controls.
Results: Despite increased lesion complexity in the TAXUS Liberté group, neointimal formation at 9 months was similar in both stents (TAXUS Liberté 13.8±11.0%; TAXUS Express 13.1±13.8%, P=.56). However, this neointima covered more of the overall stent in the TAXUS Liberté (67.9±32.5%) compared with the TAXUS Express (54.4±37.2%, P<.001), suggesting more uniform neointimal distribution. TAXUS Liberté also showed less pronounced negative remodeling at both stent edges and had significantly less (4.3% vs. 9.6%, P=.04) late incomplete stent apposition (ISA).
Conclusions: Despite identical polymer and drug release characteristic, the thin-strut TAXUS Liberté stent demonstrates improved neointimal coverage, better edge remodeling, and less late ISA vs. TAXUS Express, hereby highlighting the importance of the platform design for drug-eluting stents.