Kinematics of cervical total disc replacement adjacent to a two-level, straight versus lordotic fusion.

Journal: Spine
Published:
Abstract

Methods: In vitro biomechanical study. Objective: To characterize cervical total disc replacement (TDR) kinematics above two-level fusion, and to determine the effect of fusion alignment on TDR response.

Background: Cervical TDR may be a promising alternative for a symptomatic adjacent level after prior multilevel cervical fusion. However, little is known about the TDR kinematics in this setting.

Methods: Eight human cadaveric cervical spines (C2-T1, age: 59 ± 8.6 years) were tested intact, after simulated two-level fusion (C4-C6) in lordotic alignment and then in straight alignment, and after C3-C4 TDR above the C4-C6 fusion in lordotic and straight alignments. Fusion was simulated using an external fixator apparatus, allowing easy adjustment of C4-C6 fusion alignment, and restoration to intact state upon disassembly. Specimens were tested in flexion-extension using hybrid testing protocols.

Results: The external fixator device significantly reduced range of motion (ROM) at C4-C6 to 2.0 ± 0.6°, a reduction of 89 ± 3.0% (P < 0.05). Removal of the fusion construct restored the motion response of the spinal segments to their intact state. The C3-C4 TDR resulted in less motion as compared to the intact segment when the disc prosthesis was implanted either as a stand-alone procedure or above a two-level fusion. The decrease in motion of C3-C4 TDR was significant for both lordotic and straight fusions across C4-C6 (P < 0.05). Flexion and extension moments needed to bring the cervical spine to similar C2 motion endpoints significantly increased for the TDR above a two-level fusion compared to TDR alone (P < 0.05). Lordotic fusion required significantly greater flexion moment, whereas straight fusion required significantly greater extension moment (P < 0.05).

Conclusions: TDR placed adjacent to a two-level fusion is subjected to a more challenging biomechanical environment as compared to a stand-alone TDR. An artificial disc used in such a clinical scenario should be able to accommodate the increased moment loads without causing impingement of its endplates or undue wear during the expected life of the prosthesis.

Authors
Shelden Martin, Alexander Ghanayem, Michael Tzermiadianos, Leonard Voronov, Robert Havey, Susan Renner, Gerard Carandang, Celeste Abjornson, Avinash Patwardhan
Relevant Conditions

Spinal Fusion