Simultaneous determination of adenine nucleotides, creatine phosphate and creatine in rat liver by high performance liquid chromatography-electrospray ionization-tandem mass spectrometry.

Journal: Journal Of Pharmaceutical And Biomedical Analysis
Published:
Abstract

A high performance liquid chromatography-electrospray ionization-tandem mass spectrometric method (HPLC-ESI-MS/MS) was developed for simultaneous determination of adenosine 5'-triphosphate (ATP), adenosine 5'-diphosphate (ADP), adenosine 5'-monophosphate (AMP), creatine phosphate (CP), and creatine in rat liver. After extraction with pre-cooled (4 °C) methanol/water (1:1, v/v), the analytes were separated on a porous graphitic carbon (Hypercarb) column (2.1 mm × 150 mm, 5 μm) using a programmed gradient elution with a mobile phase consisting of 2 mmol/L ammonium acetate in water and 2 mmol/L ammonium acetate in acetonitrile (pH=10.0). The analytes were detected in a way of multiple reaction monitoring (MRM) under negative scan mode by a triple quadrupole mass spectrometer with electrospray ionization (ESI). An external calibration method with linear ranges from 10 to 5000 ng/mL for the five target compounds was used for quantification with a correlation coefficients≥0.9973. The limits of detection and limits of quantification for all analytes were in ranges from 0.50 to 1.5 ng/mL and 1.6 to 0.5 ng/mL, respectively. The average recoveries spiked in three levels were from 77.2% to 102% and precisions expressed in RSDs were from 0.2% to 4.8%. The established method was successfully applied to determination of ATP, ADP, AMP, CP and creatine in liver tissue.

Authors
Yang Jiang, Chengjun Sun, Xueqin Ding, Ding Yuan, Kefei Chen, Bo Gao, Yi Chen, Aimin Sun