A rapid and sensitive LC/ESI-MS/MS method for quantitative analysis of docetaxel in human plasma and its application to a pharmacokinetic study.

Journal: Journal Of Chromatography. B, Analytical Technologies In The Biomedical And Life Sciences
Published:
Abstract

Docetaxel is a taxane family antineoplastic agent widely employed in cancer chemotherapy. We developed a liquid chromatography/tandem mass spectrometry method for the determination of docetaxel in human plasma. Plasma samples were deproteinized by acetonitrile containing internal standard paclitaxel. Chromatographic separation was performed on a TSKgel ODS-100 V 3 μm (50 mm × 2.0 mm i.d.) column using a mobile phase composed of acetonitrile-methanol-water-formic acid (50:5:45:0.1, v/v/v/v). Detection was performed on a triple-quadrupole tandem mass spectrometer with multiple reaction monitoring (MRM) mode via electrospray ionization (ESI) source. This method covered a linearity range of 5-5000 ng/mL with the lower limit of quantification of 5 ng/mL. The intra-day precision and inter-day precision (R.S.D.) of analysis were less than 6.7%, and the accuracy (R.E.) was within ± 9.0% at the concentrations of 5, 20, 200, and 2000 ng/mL. The total run time was 5.0 min. This method was successfully applied for clinical pharmacokinetic investigation.

Authors
Hiroaki Yamaguchi, Asuka Fujikawa, Hajime Ito, Nobuaki Tanaka, Ayako Furugen, Kazuaki Miyamori, Natsuko Takahashi, Jiro Ogura, Masaki Kobayashi, Takehiro Yamada, Nariyasu Mano, Ken Iseki