Intrapartum heart rate ambiguity: a comparison of cardiotocogram and abdominal fetal electrocardiogram with maternal electrocardiogram.

Journal: Gynecologic And Obstetric Investigation
Published:
Abstract

Objective: To investigate the presence of signal ambiguity of intrapartum fetal heart rate (FHR) monitoring during delivery by comparing simultaneous cardiotocogram (CTG), abdominal fetal electrocardiogram (ECG) with continuous maternal ECG.

Methods: A total of 144 simultaneous CTG (Corometrics 250 series), abdominal fetal ECG (Monica -AN24™) and maternal ECG (Monica AN24™) recordings were evaluated. Methods: When the FHR is within 5 bpm of the maternal heart rate (MHR) acquired from the ECG, it is classified as 'MHR/FHR ambiguity'. Statistical analyses were performed with Fisher's exact test and the Wilcoxon signed-rank test.

Results: Comparison of abdominal fetal ECG against CTG demonstrates significantly less 'MHR/FHR ambiguity' in both the first stage (mean 0.70 vs. 1.22%, p < 0.001) and second stage of labour (mean 3.30 vs. 6.20%, p < 0.001).

Conclusions: Intrapartum FHR monitoring in daily practice via the CTG modality provides significantly more 'MHR/FHR ambiguity' than abdominal fetal ECG, which also provides additional information on the MHR.

Authors
Joscha Reinhard, Barrie Hayes Gill, Sven Schiermeier, Hendrike Hatzmann, Tomas Heinrich, Frank Louwen