Robot-assisted aortic valve replacement using a novel sutureless bovine pericardial prosthesis: proof of concept as an alternative to percutaneous implantation.
Objective: : Percutaneous aortic valve implantation within native valve calcium has progressed to clinical use despite the absence of data proving equivalence to complete surgical excision and prosthetic valve replacement. A novel self-expanding sutureless bovine pericardial prosthesis (Sorin Perceval) derived from a proven stented valve has been successfully used in humans recently through an open transaortic approach. We sought to develop a minimally invasive technique for native aortic valve excision and sutureless prosthetic aortic valve replacement using robot assistance.
Methods: : The da Vinci S-HD system was used to open and suspend the pericardium anterior to the phrenic nerve in cadavers. A transthoracic cross-clamp was placed across the midascending aorta, following which a transverse aortotomy was made. The native aortic valve cusps were excised, and annular calcium was removed with robotic instruments. After placement of three guide sutures, the Perceval self-expanding pericardial prosthesis mounted on a flexible delivery system was inserted through a working port and lowered into the aortic annulus.
Results: : Successful implantation of all valves was possible using a 3-cm right second intercostal space working port, along with two additional 1-cm instrument ports. A standard transverse aortotomy was sufficient for examination/debridement of the native aortic valve cusps, sizing of the annulus, and deployment of the nitinol-stented, bovine pericardial prosthesis. Delivery, seating, and stability of the device were easily confirmed above and below the aortic valve annulus using the robotic camera.
Conclusions: : Complete excision of diseased native aortic valve cusps with robot assistance facilitates accurate and reproducible aortic valve replacement using a novel self-expanding sutureless version of a proven bovine pericardial prosthesis. This approach is comparable to the current surgical gold standard and is ready for clinical use as an alternative to percutaneous aortic valve implantation.