LC-MS/MS determination of apixaban (BMS-562247) and its major metabolite in human plasma: an application of polarity switching and monolithic HPLC column.

Journal: Bioanalysis
Published:
Abstract

Background: apixaban (BMS-562247) (Eliquis(®)) is a novel, orally active, selective, direct, reversible inhibitor of the coagulation factor Xa (FXa). A sensitive and reliable method was developed and validated for the measurement of apixaban (BMS-562247) and its major circulating metabolite (BMS-730823) in human citrated plasma for use in clinical testing.

Results: A 0.100 ml portion of citrated plasma sample was extracted and analyzed by LC-MS/MS. Run times were approximately 3 min. The lower limit of quantification (LLOQ) was 1.00 ng/ml for BMS-562247 and 5.00 ng/ml for BMS-730823. Intra- and inter-assay precision values for replicate QC control samples were within ≤5.36% for both analytes (≤7.52% at the LLOQ). The accuracy for both analytes was within ±9.00%.

Conclusions: The method was demonstrated to be sensitive, selective and robust, and was successfully used to support clinical studies.

Authors
Janice Pursley, Jim Shen, Alan Schuster, Oanh Dang, Jim Lehman, Michael Buonarati, Yan Song, Anne-francoise Aubry, Mark Arnold