How does ankle-foot orthosis stiffness affect gait in patients with lower limb salvage?

Journal: Clinical Orthopaedics And Related Research
Published:
Abstract

Background: Ankle-foot orthoses (AFOs) are commonly prescribed during rehabilitation after limb salvage. AFO stiffness is selected to help mitigate gait deficiencies. A new custom dynamic AFO, the Intrepid Dynamic Exoskeletal Orthosis (IDEO), is available to injured service members but prescription guidelines are limited.

Objective: In this study we ask (1) does dynamic AFO stiffness affect gait parameters such as joint angles, moments, and powers; and (2) can a given dynamic AFO stiffness normalize gait mechanics to noninjured control subjects?

Methods: Thirteen patients with lower limb salvage (ankle arthrodesis, neuropathy, foot/ankle reconstruction, etc) after major lower extremity trauma and 13 control subjects who had no lower extremity trauma and wore no orthosis underwent gait analysis at a standardized speed. Patients wore their custom IDEO with posterior struts of three different stiffnesses: nominal (clinically prescribed stiffness), compliant (20% less stiff), and stiff (20% stiffer). Joint angles, moments, powers, and ground reaction forces were compared across the varying stiffnesses of the orthoses tested and between the patient and control groups.

Results: An increase in AFO compliance resulted in 20% to 26% less knee flexion relative to the nominal (p = 0.003) and stiff (p = 0.001) conditions, respectively. Ankle range of motion and power generation were, on average, 56% (p < 0.001) and 63% (p < 0.001), respectively, less than controls as a result of the relatively fixed ankle position.

Conclusions: Patients with limb salvage readily adapted to different dynamic AFO stiffnesses and demonstrated few biomechanical differences among conditions during walking. None of the stiffness conditions normalized gait to controls. Conclusions: The general lack of differences across a 40% range of strut stiffness suggests that orthotists do not need to invest large amounts of time identifying optimal device stiffness for patients who use dynamic AFOs for low-impact activities such as walking. However, choosing a stiffer strut may more readily translate to higher-impact activities and offer less chance of mechanical failure.

Authors
Elizabeth Russell Esposito, Ryan Blanck, Nicole Harper, Joseph Hsu, Jason Wilken