Hemodynamic effects of left atrial or left ventricular cannulation for acute circulatory support in a bovine model of left heart injury.

Journal: ASAIO Journal (American Society For Artificial Internal Organs : 1992)
Published:
Abstract

Our objective was to examine the hemodynamic effects of a trans-aortic axial flow catheter (Impella CP) in the left ventricle (LV) versus left atrial (LA) to femoral artery bypass using a centrifugal pump (TandemHeart: TH) in a bovine model of acute LV injury. In three male calves, we performed sequential activation of a CP then TH device in each animal. After 60 minutes of left anterior descending artery ligation, a CP was activated at maximal power. The CP was then removed and the TH activated at 5,500 then a maximum of 7,500 rotations per minute (RPM). The CP generated a maximum 3.1 ± 0.2 L/minute (LPM) of flow, whereas the TH at 5,500 and 7,500 RPM generated 3.1 ± 0.4 and 4.4 ± 0.3 LPM. At 3.1 LPM, the CP and TH reduced LV stroke work (LVSW) similarly. The TH reduced stroke volume, whereas the CP did not. The CP reduced end-systolic pressure, whereas the TH did not. At a maximum flow of 4.4 LPM, the TH provided a greater reduction in LVSW than maximal CP activation. This is the first report to compare the hemodynamic effects of trans-aortic LV unloading versus LA-to-femoral artery (FA) bypass.

Authors
Navin Kapur, Vikram Paruchuri, Duc Pham, Lara Reyelt, Barbara Murphy, Corinna Beale, Courtney Bogins, Daniel Wiener, James Nilson, Michele Esposito, Scott Perkins, George Perides, Richard Karas