How sensitive is the deltoid moment arm to humeral offset changes with reverse total shoulder arthroplasty?
Background: Reverse total shoulder arthroplasty commonly treats cuff-deficient or osteoarthritic shoulders not amenable to rotator cuff repair. This study investigates deltoid moment arm sensitivity to variations in the joint center and humeral offset of 3 representative reverse total shoulder arthroplasty subjects. We hypothesized that a superior joint implant placement may exist, indicated by muscle moment arms, compared with the current actual surgical implant configuration.
Methods: Moment arms for the anterior, lateral, and posterior aspects of the deltoid muscle were determined for 1521 perturbations of the humeral offset location away from the surgical placement in a subject-specific musculoskeletal model with motion defined by subject-specific in vivo abduction kinematics. The humeral offset was varied from its surgical position ±4 mm in the anterior/posterior direction, ±12 mm in the medial/lateral direction, and -10 to 14 mm in the superior/inferior direction.
Results: The anterior deltoid moment arm varied in humeral offset and center of rotation up to 20 mm, primarily in the medial/lateral and superior/inferior directions. The lateral deltoid moment arm varied in humeral offset up to 20 mm, primarily in the medial/lateral and anterior/posterior directions. The posterior deltoid moment arm varied up to 15 mm, primarily in early abduction, and was most sensitive to humeral offset changes in the superior/inferior direction.
Conclusions: High variations in muscle moment arms were found for all 3 deltoid components, presenting an opportunity to dramatically change the deltoid moment arms through surgical placement of the reverse shoulder components and by varying the overall offset of the humerus. Methods: Basic Science Study; Computer Modeling.