On the limitations of partial Fourier acquisition in phase-contrast MRI of turbulent kinetic energy.
Purpose: To investigate limitations of partial Fourier acquisition in phase-contrast MRI of turbulent kinetic energy (TKE).
Methods: To assess the validity of partial Fourier reconstruction of TKE and phase images, computational fluid dynamics data of mean and turbulent velocities in a stenotic U-bend phantom was used. Partial Fourier acquisition with 75% k-space coverage was simulated and TKE data were reconstructed using zero-filling, homodyne reconstruction, and the method of projections onto convex sets (POCS). Results were compared to data from fully sampled k-space and 75% symmetric sampling. In addition, compressed sensing (CS) reconstruction was compared for a standard variable density sampling pattern and a variable density sampling pattern combined with 75% partial Fourier. For illustration purposes, in vivo examples of velocity magnitude and TKE maps of aortic flow reconstructed with the different methods are provided.
Results: In accordance with theory, partial Fourier reconstruction of TKE maps from phase-contrast data results in artifacts relative to fully sampled data. It is demonstrated that neither homodyne reconstruction nor POCS can improve reconstruction of TKE data with respect to zero-filling reconstruction when compared to ground-truth (RMS error: 4.70%, 4.34%, and 2.45% for homodyne, POCS, and zero-filling reconstruction of in vivo data, respectively). CS reconstruction from data acquired with partial Fourier did not recover the resolution loss incurred by partial Fourier sampling.
Conclusion: Partial Fourier reconstruction of TKE maps from phase-contrast data does not yield a benefit over zero-filling reconstruction. In consequence, symmetric sampling is preferred over partial Fourier acquisition for a given number of phase-encodes in phase-contrast MRI.