Ultrasound carpal tunnel syndrome: additional criteria for diagnosis.
Aim: To define the usefulness of the cross-sectional area (CSA) of the median nerve distal to the carpal tunnel in addition to other established common parameters in the diagnosis of carpal tunnel syndrome (CTS). Materials and
Methods: Forty-four wrists from 24 symptomatic CTS patients and 32 wrists from 17 asymptomatic volunteers were evaluated by ultrasound. The CSA of the median nerve was measured at four pre-selected levels, i.e., proximal, inlet, outlet, and distal to the carpal tunnel. The flattening ratio, intraneural vascularity, neural fasciculation, and retinacular palmar bowing were also assessed.
Results: Significant differences were found between the CTS and control groups for median nerve CSA proximal and distal (p<0.001) to the tunnel as well as retinacular bowing (p<0.001). Using the receiver operating characteristic (ROC) curve, the sensitivity, specificity, and accuracy of using a cut-off of >14 mm2 of CSA proximal and distal to the tunnel were 75%, 87.5%, 86.8% and 63.6%, 100%, 78.9%, respectively. Using either CSA proximal or distal to the tunnel or bowing retinaculum at the outlet >1 mm yielded a sensitivity, specificity, and accuracy of 100%, 84.3% and 93.4%, respectively.
Conclusion: The median nerve CSA proximal and distal to the carpal tunnel and bowing of the retinaculum at the outlet are helpful in diagnosis of CTS.