Image quality of virtual monochromatic images obtained using 320-detector row CT: A phantom study evaluating the effects of iterative reconstruction and body size.
Objective: To compare the image quality between virtual monochromatic spectral (VMS) images obtained using 320-row detector CT and polychromatic 120-kVp images reconstructed with or without iterative reconstruction using various phantom sizes.
Methods: Torso phantoms simulating three patient sizes and containing four syringes filled with water or different contrast media (5, 10, 15mgI/mL15mgI/ml) were used. The phantoms were scanned using dual-energy (80/135-kVp) and single-energy (120-kVp) protocols at different settings (20mGy, 12mGy, and 6mGy). VMS images were generated at 1-keV intervals (range, 35-135keV). Both the VMS images and the single-energy 120-kVp images were reconstructed using filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR-3D). The signal-to-noise ratio (SNR), and the contrast-to-noise ratio (CNR) were assessed.
Results: Using FBP reconstruction, the SNR and CNR of the VMS images were lower than or similar to those of the 120-kVp images for most dose settings. Using AIDR-3D reconstruction, however, the 70-keV VMS images had higher SNRs and CNRs than the 120-kVp images at most settings.
Conclusions: The image quality of VMS images with FBP reconstruction tended to be lower than that of the 120-kVp images. With the use of AIDR-3D, however, approximately 70-keV VMS images had a higher image quality than the 120-kVp images.