Cortical pedicle screw placement in lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study.

Journal: Journal Of Orthopaedic Science : Official Journal Of The Japanese Orthopaedic Association
Published:
Abstract

Background: Cortical pedicle screw placement is an attractive technique in terms of both fixation strength and less invasiveness. However, to insert the screw with penetrating cortical bone on the ideal trajectory is technically demanding. The use of three-dimensional (3D) patient-matched guides may facilitate the use of this technique.

Objective: To examine the accuracy of cortical screw placement using a patient-matched targeting guide with a cadaveric study assessing the accuracy.

Methods: The 3D planning of the pedicle screw placement, including the location at which the screw would pass through the center of the pedicle, sagittal/transverse trajectory (angle), length, and diameter, was developed using 3D CAD design software. Three-dimensional guides based on the preoperative planning were created for three cadaveric specimens (L1 to S1, 36 pedicles). Screws (n = 18) and pins (n = 18) were placed using K-wire or drill-based guides, without X-ray exposure. Actual positioning was compared to the preoperative plan by superimposing the inserted screws/pins based on postoperative CT. The placement accuracy was graded based on the degree of perforation of the pedicle by the pedicle screw or pin using an acceptance criterion (no perforation; Grade A, 0-2 mm; Grade B, 2-4 mm; and Grade C, >4 mm). The mean deviation between the planned and inserted screw positions on the coronal plane at the midpoint of the pedicle was compared to the accuracy of screw guide for traditional pedicle screw trajectory (0.70 mm).

Results: Of 35 evaluated screws and pins, 32 (91.4%) were inserted completely inside the pedicle. All pedicle perforation was within 2 mm. The mean deviation from the plan at the midpoint of pedicle was 0.66 mm; thus, the accuracy was within the predefined criteria.

Conclusions: Cortical pedicle screw placement using 3D-patient matched guides is accurate. Further clinical studies are required to confirm the radiographic and clinical effects.