Analysis of Environmental Protection Agency priority endocrine disruptor hormones and bisphenol A in tap, surface and wastewater by online concentration liquid chromatography tandem mass spectrometry.

Journal: Journal Of Chromatography. A
Published:
Abstract

The list of endocrine disrupting compounds (EDCs) defined under U.S. EPA Method 539 was recently expanded to include additional hormones and bisphenol A (BPA). Here, we validated a fast and robust alternative method compliant with Method 539.1 requirements in diverse water matrixes (i.e., ultra-pure water, tap water, surface water, and wastewater influent and effluent). Automated large volume injection solid phase extraction (SPE) coupled on-line to ultra-high-performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) was investigated for this purpose. The surveyed molecules included 13 EPA-priority hormones (testosterone, progesterone, medroxyprogesterone, levonorgestrel, norethindrone, androstenedione, estrone, β-estradiol, α-estradiol, equilin, equilenin, ethinylestradiol, estriol) and BPA. Combinations of ionization source and mobile phases were optimized for improved sensitivity. Suitable chromatographic performances were obtained and the implementation of an on-line SPE washing step consecutive to sample loading was investigated. On-line SPE extraction efficiencies in acceptable ranges (64-79%) and detection limits in the order of nanogram per liter or sub-nanogram per liter were obtained. The linearity range extended over 2-3 orders of magnitude, with determination coefficients (R2) typically > 0.9980. Robust precision and trueness complying with acceptance criteria (70-130%) were obtained for the scope of analytes/matrix combinations. Limited internal standard variations were also observed across samples (±18%), well within the ±50% acceptance criterion. The method was successfully applied to field-collected samples in Canada and summed EDC concentrations were reported in the range of 0.80-2.8 ng L-1, 6.8-19 ng L-1, 260-790 ng L-1, and 37-360 ng L-1 in tap water, surface water, effluent and influent wastewater samples, respectively.

Authors
Ken Goeury, Sung Vo Duy, Gabriel Munoz, Michèle Prévost, Sébastien Sauvé