Comparison of Less Invasive Stabilization System Plate and Retrograde Intramedullary Nail in the Fixation of Femoral Supracondylar Fractures in the Elderly: A Biomechanical Study.
Objective: To compare the biomechanical stabilities of less invasive stabilization system (LISS) plate and retrograde intramedullary nail (IMN) for the comminuted femoral supracondylar fracture fractures in the elderly.
Methods: Sixteen pairs of embalmed cadaver femurs were obtained to simulate a comminuted supracondylar femur fracture (AO/OTA33-A3) gap model. All left-side specimens were fixed with LISS plate, and retrograde IMN were applied to the right-side specimens. All specimens were tested in torsional, axial and cyclic load mode on an Instron testing machine.
Results: The mean torsional stiffness for LISS plate group was 34.1% greater than retrograde IMN group (2.90 vs. 1.91 Nm/degree, P = 0.002), but the mean axial stiffness was greater for the retrograde IMN (199.16 vs. 303.93 N/mm, P < 0.001). The total deformation of LISS plate caused by cyclic axial loading was greater than retrograde IMN (4.17 vs. 3.57 mm, P = 0.014). Significantly less mean irreversible deformation was detected in LISS plate than in retrograde IMN (1.64 vs. 1.69 mm, P = 0.699). Failure loads of the constructs were significantly different between the two groups (LISS plate: 2941±128 N; retrograde IMN: 4022±176 N, P < 0.001).
Conclusions: For comminuted femoral supracondylar fractures in the elderly, the tested instruments can both maintain sufficient biomechanical stabilities, but retrograde IMN is superior to LISS plate in deformation of fracture site.