Mechanical Study of the Properties of Sutures used in Orthopedics Surgeries.

Journal: Revista Brasileira De Ortopedia
Published:
Abstract

Objective  To evaluate the mechanical properties of sutures commonly used in orthopedic surgeries, characterizing their behavior through tensile tests and determining which one has greater mechanical strength. Method  Tensile tests of different sutures were performed in a mechanical testing machine BME 10 kN, using a 50 kgf maximum capacity loading cell. Seven samples from each suture material were tested. Both ends of the sample material were fixed in the proper metal claw, maintaining an initial length of 5 cm. Tests were performed at a speed of 20 mm/minute and at room temperature, recording data for maximum displacement and maximum force at the rupture point. Results  FiberWire® #2 (Artrhex, Naples, FL, USA) presented the highest mean strength of rupture (240.17 N), followed by HiFi® #2 (Conmed, Utica, NY, USA) (213.39N) and Ethibond® #5 (Ethicon Inc., Somerville, NJ, USA) (207.38 N). Ethibond® #2 (Ethicon Inc., Somerville, NJ, USA) had the lowest mean strength of rupture (97.8 N). Conclusion  Non-absorbable braided polyblend sutures, more recently introduced, are superior to conventional, braided polyester sutures, and FiberWire® #2 is the most resistant suture evaluated in the present study.

Authors
Leandro Gomide, Dagoberto De Campos, Cleudmar Araújo, Gabriela Menegaz, Rafael Cardoso, Sérgio Saad Júnior

Similar Publications