Effects of assist parameter on the performance of proportional assist ventilation in a lung model of chronic obstructive pulmonary disease.

Journal: Respiratory Medicine And Research
Published:
Abstract

Background: How the assist parameters affect synchronization and inspiratory workload in proportional assist ventilation (PAV) remains unknown.

Objective: This bench study aimed to optimize the PAV parameters by evaluating their effects on patient-ventilator synchrony and work of breathing (WOB) in a chronic obstructive pulmonary disease (COPD) model during noninvasive ventilation, compared with the pressure support ventilation (PSV) mode.

Methods: The Respironics V60 ventilator was connected to an ASL5000 lung simulator, which simulates lung mechanics in COPD (compliance, 50mL/cmH2O; expiratory resistance, 20 cmH2O/L/s; respiratory rate, 15 breaths/min; inspiratory time, 1.6 s). PAV was applied with different assistance levels, including flow assist (FA, 40-90% respiratory resistance) and volume assist (VA, 50-90% elastance). PSV was assessed using the same model. Measurements were obtained at a leak flow rate of 25-28 L/min. Performance characteristics, simulator-ventilator synchrony, and WOB were assessed.

Results: Runaway was prone to occur, and severe premature cycling was observed with VA75+FA level>65%. Compared with PSV, lower tidal volume (≤400mL) was observed during PAV with VA75+FA40-50 and FA50+VA40-80; similar and improved cycling synchrony was observed for FA50+VA80 and FA50+VA90 (cycling delay: -117.60±6.13 and -61.50±8.03 vs. -101.20±7.32ms). The reduced triggering workload was observed for VA75+FA60-65 and FA50+VA80-90. Total and patient WOB was improved with all tested assist level combinations, except for FA50+VA90.

Conclusions: PAV reduces WOB but can induce asynchrony if improper settings are set, but the most optimal settings still need more clinical observations.

Authors
Y Chen, Y Yuan, C Cai, F Li, X Zhou