A Dose Simulation X-Ray Software: An Innovating Tool to Reduce Chest Radiograph Exposure in Children.

Journal: Journal Of Thoracic Imaging
Published:
Abstract

Objective: Chest radiography is one of the most frequent x-ray examinations performed on children. Reducing the delivered dose is always a major task. The objective of our study was to determine the minimum dose to be delivered while maintaining the image quality of chest radiographs, using dose reduction simulation software.

Methods: We included 60 children who had had a chest radiography in 5 groups established according to the diagnostic reference levels equitably represented by weight ranges. The software simulated for each radiograph 6 additional simulated photonic noise images corresponding to 100%, 80%, 64%, 50%, 40%, and 32% of the initial dose. The 360 radiographs were blindly scored by 2 radiologists, according to the 7 European quality criteria and a subjective criterion of interpretability, using a semiquantitative visual Lickert scale.

Results: There was no significant difference in scoring between the reference radiograph (100%) and simulated radiographs at 80% of the dose in children between 5 and 20 kg, 50% of the dose in children between 20 and 30 kg, and between simulated radiographs at 40% of the dose in children over 30 kg. Interobserver reproducibility was moderate to excellent.

Conclusions: Chest radiography dose might be reduced by 20% in children between 5 and 20 kg, 50% in children between 20 and 30 kg, and 60% in children over 30 kg, without any difference in the image quality appreciation. Software that produced simulated x-ray with decreasing delivered dose is an innovating tool for an optimization process.

Authors
Benjamin Meunier, Ronan Boënnec, Paul-armand Dujardin, Jean Rafin, Dominique Sirinelli, Guillaume Chassagnon, Baptiste Morel