Low-dose CT of paediatric paranasal sinus using an ultra-low tube voltage (70 kVp) combined with the flash technique.
Aim: To evaluate the radiation dose and diagnostic image quality of low-dose computed tomography (CT) of the paranasal sinus in children, with acquisition at an ultra-low tube voltage (70 kVp) combined with the Flash technique. Materials and
Methods: Eighty paediatric patients underwent CT of the paranasal sinus and were divided into two groups according to different protocols (group A: 80 kVp protocol with conventional spiral mode [n=40] and group B: 70 kVp protocol with Flash scan mode [n=40]). For each examination, the CT dose index (CTDIvol), dose-length product (DLP), and effective dose (ED) were estimated. The image noise, signal-to-noise ratio (SNR), and overall subjective diagnostic image quality were also evaluated.
Results: For radiation dose, the CTDIvol (mGy), DLP (mGy·cm), and ED (mSv) values of the 70 kVp protocol were significantly lower than those of the 80 kVp protocol (CTDIvol: 1.57±0.009 versus 0.39±0.004 mGy, p<0.001; DLP: 19.88±2.01 versus 6.31±0.52 mGy·cm, p<0.001; ED: 0.079±0.016 versus 0.024±0.005 mSv, p<0.001). Compared with those of the 80-kVp protocol, the image noise increased by 40.7% (p=0.113), the SNRsoft-tissue decreased by 48.9%, and the SNRbone increased by 10.1% with the 70-kVp protocol (p=0.176 and 0.227, respectively). There was no significant difference in the overall subjective image quality grades between these two groups (p=0.15).
Conclusion: When imaging the paranasal sinus in children, an ultra-low tube voltage (70 kVp) combined with the Flash CT technique can reduce the radiation dose significantly while maintaining diagnostic image quality with clinically acceptable image noise.