Evaluation of identification accuracy using AIQS for GC-MS for measuring heavily contaminated samples.

Journal: Chemosphere
Published:
Abstract

Automated identification and quantification systems with gas chromatography-mass spectrometry (GC-MS) (i.e., AIQS-GC) are used as a simple and comprehensive method for screening chemicals existing in the environment and are expected to be useful for emergency surveys in the event of a disaster. However, reports on the potential of AIQS-GC in heavily contaminated samples (HCSs) are limited. In this study, the identification performance of AIQS-GC was confirmed by comparing the exact mass of the targets identified by AIQS-GC with the measured accurate mass using GC-quadrupole-time-of-flight MS (GC-QTofMS) and by employing firefighting wastewater as HCS. In HCS, the mass spectrum interference was determined to cause false positives. The GC-QTofMS method revealed the presence of false positives and the false rate of AIQS-GC in HCS. Herein, AIQS-GC showed high identification accuracy in a normal sample such as river water. Conversely, in HCS, AIQS-GC may lead to incorrect evaluations. The combination of AIQS-GC and support method using GC-QTofMS, which can avoid the false positive is extremely useful for the rapid and easy analysis of HCS.