Magnetic covalent-organic frameworks for the simultaneous extraction of eleven emerging aromatic disinfection byproducts in water samples coupled with UHPLC-MS/MS determination.

Journal: Journal Of Hazardous Materials
Published:
Abstract

A simple method based on magnetic solid-phase extraction (MSPE) was developed for the simultaneous extraction of eleven emerging aromatic disinfection byproducts (DBPs) in water samples coupled with ultrahigh-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) determination. A magnetic covalent-organic framework (COF) material, namely, Fe3O4 @TpBD, was facilely synthesized and fully characterized, followed by an MSPE process. Several important MSPE parameters, such as the magnetic ratio, Fe3O4 @TpBD amount and sample pH, were systematically investigated. Under optimal conditions, the limits of detection and quantification of this COF-MSPE-UHPLC-MS/MS method were as low as 0.07-1.81 ng/L and 0.24-5.99 ng/L, respectively. Good precision was obtained with relative standard deviations (RSDs) of 1.3-10.9% (intraday) and 4.3-15.9% (interday). Furthermore, the validated method was proven applicable to real water samples; for example, the recoveries were 86.8-115.1% for the secondary effluent, and several DBPs in swimming pool water were detected. Notably, the MSPE process required only 7 min, ensuring that the DBPs were relatively stable during the whole analysis process and that Fe3O4 @TpBD demonstrated excellent reusability. The COF-based MSPE method with simplicity, rapidity and efficiency provided an ideal sample pretreatment alternative to determine trace DBPs in complex matrices.

Authors
Shuang Li, Jiping Ma, Gege Wu, Jinhua Li, Xiaoyan Wang, Lingxin Chen