Abstract

The purpose of this study was to assess the clinical performance of STANDARD F COVID-19 Ag FIA (SD Biosensor Inc., Gyeonggi-do, Republic of Korea), a rapid antigen detection test (RADT) for diagnosing SARS-CoV-2, in patients attended at the Emergency Department with signs or symptoms compatible with COVID-19 that had started in the last 5 days. The clinical performance of the antigen test was compared with RT-PCR, the reference standard. We included 663 specimens from non-repetitive patients. Clinical sensitivity and specificity were 84.0% (95% CI 76.1-89.7) and 99.6% (95% CI 98.5-99.9), respectively. The positive and negative predictive values were 98.1% (95% CI 92.7-99.7) and 96.4% (95% CI 94.4-97.7), respectively. The kappa index agreement between RT-PCR and the RADT was 0.89 (95% CI 0.84-0.93). We concluded that STANDARD F COVID-19 Ag FIA is an excellent first-line RADT method to diagnose symptomatic patients in the emergency department.

Authors
Sergio García Fernández, Daniel Pablo Marcos, Silvia De La Fuente, María José Rodríguez, Mónica Gozalo, Jesús Rodríguez Lozano, José Méndez Legaza, Jorge Calvo