Comparison of stent displacement and displacement force after endovascular aneurysm repair with cross-limb or parallel-limb stent

Journal: Sheng Wu Yi Xue Gong Cheng Xue Za Zhi = Journal Of Biomedical Engineering = Shengwu Yixue Gongchengxue Zazhi
Abstract

This study aims to investigate whether displacement force on stents can accurately represents the displacement of the stent after endovascular aneurysm repair (EVAR) by comparing the measured stent displacement with the displacement forces calculated by computational fluid dynamics (CFD). And the effect of cross-limb and parallel-limb EVAR on stent displacements is further studied. Based on our objective, in this study, ten cross-limb EVAR patients and ten parallel-limb EVAR patients in West China Hospital of Sichuan University were enrolled. Patient-specific models were first reconstructed based on the computed tomography angiography images, then the stent displacements were measured, and the displacement forces acting on the stents were calculated by CFD. Finally, the