Comparative analysis between STANDARD-E Covi-FERON ELISA with pre-existing IFN-γ release assays and determination of the optimum cutoff value for assessment of T-Cell response to SARS-CoV-2.
Background: Interferon-gamma (IFN-γ) release assays (IGRAs) are useful for the assessment of the T-cell response to severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2). We aimed to assess the performance of the newly developed IGRA ELISA test compared to the pre-existing assays and to validate the cutoff value in real-world conditions.
Methods: We enrolled 219 participants and assessed agreement between STANDARD-E Covi-FERON ELISA with Quanti-FERON SARS-CoV-2 (QFN SARS-CoV-2), as well as with T SPOT Discovery SARS-CoV-2 based on Cohen's kappa-index. We further determined the optimal cutoff value for the Covi-FERON ELISA according to the immune response to vaccinations or infections.
Results: We found a moderate agreement between Covi-FERON ELISA and QFN SARS-CoV-2 before vaccination (kappa-index = 0.71), whereas a weak agreement after the first (kappa-index = 0.40) and second vaccinations (kappa-index = 0.46). However, the analysis between Covi-FERON ELISA and T SPOT assay demonstrated a strong agreement (kappa-index >0.7). The cut-off value of the OS (original spike) marker was 0.759 IU/mL with a sensitivity of 96.3% and specificity of 78.7%, and that of the variant spike (VS) marker was 0.663 IU/mL with a sensitivity and specificity of 77.8% and 80.6%, respectively.
Conclusions: The newly determined cut-off value may provide an optimum value to minimize and prevent the occurrence of false-negative or false-positive during the assessment of T-cell immune response using Covi-FERON ELISA under real-world conditions.