Feasibility evaluation of novel AI-based deep-learning contouring algorithm for radiotherapy.
Objective: To evaluate the clinical feasibility of the Siemens Healthineers AI-Rad Companion Organs RT VA30A (Organs-RT) auto-contouring algorithm for organs at risk (OARs) of the pelvis, thorax, and head and neck (H&N).
Methods: Computed tomography (CT) datasets from 30 patients (10 pelvis, 10 thorax, and 10 H&N) were collected. Four sets of OARs were generated on each scan, one set by Organs-RT and the others by three experienced users independently. A physician (expert) then evaluated each contour by assigning a score from the following scale: 1-Must Redo, 2-Major Edits, 3-Minor Edits, 4-Clinically usable. Using the highest-scored OAR from the human users as a reference, the contours generated by Organs-RT were evaluated via Dice Similarity Coefficient (DSC), Hausdorff Distance (HDD), Mean Distance to Agreement (mDTA), Volume comparison, and visual inspection. Additionally, each human user recorded the time to delineate each structure set and time-saving efficiency was measured.
Results: The average DSC obtained for the pelvic OARs ranged between (0.81 ± 0.06)Rectum and (0.94 ± 0.03)Bladder . (0.75 ± 0.09)Esophagus to