Improvement of coronary stent visualization using ultra-high-resolution photon-counting detector CT.

Journal: European Radiology
Published:
Abstract

Objective: This study aimed to compare the image quality and diagnostic performance of standard-resolution (SR) and ultra-high-resolution (UHR) coronary CT angiography (CCTA) based on photon-counting detector CT (PCD-CT) of coronary stents and explore the best reconstruction kernel for stent imaging.

Methods: From July 2023 to September 2023, patients were enrolled to undergo CCTA using a dual-source PCD-CT system after coronary angioplasty with stent placement. SR images with a slice thickness/increment of 0.6/0.4 mm were reconstructed using a vascular kernel (Bv48), while UHR images with a slice thickness/increment of 0.2/0.2 mm were reconstructed using vascular kernels of six sharpness levels (Bv48, Bv56, Bv60, Bv64, Bv72, and Bv76). The in-stent lumen diameters were evaluated. Subjective image quality was also evaluated by a 5-point Likert scale. Invasive coronary angiography was conducted in 12 patients (25 stents).

Results: Sixty-nine patients (68.0 [61.0, 73.0] years, 46 males) with 131 stents were included. All UHR images had significantly larger in-stent lumen diameter than SR images (p < 0.001). Specifically, UHR-Bv72 and UHR-Bv76 for in-stent lumen diameter (2.17 [1.93, 2.63] mm versus 2.20 [1.93, 2.59] mm) ranked the two best kernels. The subjective analysis demonstrated that UHR-Bv72 images had the most pronounced effect on reducing blooming artifacts, showcasing in-stent lumen and stent demonstration, and diagnostic confidence (p < 0.001). Furthermore, SR and UHR-Bv72 images showed a diagnostic accuracy of 78.3% (95% confidence interval [CI]: 56.3%-92.5%) and 88.0% (95%CI: 68.8%-97.5%), respectively.

Conclusions: UHR CCTA by PCD-CT leads to significantly improved visualization and diagnostic performance of coronary stents, and Bv72 is the optimal reconstruction kernel showing the stent struts and in-stent lumen. Conclusions: The significantly improved visualization of coronary stents using ultra-high resolution CCTA could increase the diagnostic accuracy for in-stent restenosis and avoid unnecessary invasive quantitative coronary angiography, thus changing the clinical management for patients after percutaneous coronary intervention. Conclusions: Coronary stent imaging is challenging with energy-integrating detector CT due to "blooming artifacts." UHR images using a PCD-CT enhanced coronary stent visualization. UHR coronary stent imaging demonstrated improved diagnostic accuracy in clinical settings.

Authors
Le Qin, Shanshui Zhou, Haipeng Dong, Jiqiang Li, Ruiyan Zhang, Chendie Yang, Peng Liu, Zhihan Xu, Fuhua Yan, Wenjie Yang
Relevant Conditions

Coronary Heart Disease