Accuracy of Modern and Traditional Intraocular Lens Power Calculation Formulas in Pediatric Cataract Surgery.
To compare the accuracy of modern intraocular lens (IOL) power calculation formulas with that of older formulas, such as SRK/T and Hoffer Q, in pediatric cataract surgery. This retrospective study included 100 eyes of 100 children who underwent routine cataract surgery with primary IOL implantation in a bag. This study used four IOLMaster 700 integrated formulas: SRK/T, Hoffer Q, Haigis, and Barrett Universal II (BUII). In addition, the following formulas were used: EVO 2.0, Hill RBF 3.0, Hoffer QST, Kane, and PEARL DGS, which are available online. There was a statistically significant difference between SRK/T and most other formulas, except for Hoffer Q, Hoffer QST, and BUII (p < 0.05). SRK/T yielded the lowest median absolute error (MedAE) of 0.63 D. This was followed by the BUII (0.66 D), Hoffer Q, and Hoffer QST (0.68 D). SRK/T also yielded the highest percentage of cases within ± 0.50 D (43% of the cases). For patients aged 2 to 5 years, SRK/T formula yielded statistically significantly better results than all other included formulas (p < 0.05) with MedAE = 0.44 D, 58.33% and 87.50% of the cases were within ± 0.50 D and ± 1.0 D of intended refraction, respectively. The SRK/T formula showed the best IOL power calculation results in pediatric cataract surgery, followed by BUII, Hoffer Q, and Hoffer QST. In children aged 2-5 years, the SRK/T formula outperformed all other formulas, followed by the BUII and Hoffer QST formulas. In children older than 5 years, there was no statistically significant difference between the different formulas (p > 0.05); Hoffer Q and SRK/T showed slightly better MedAE in this age group (5-10 years).