Imaging near titanium total hip arthroplasty at 0.55 T compared with 3 T.

Journal: Magnetic Resonance In Medicine
Published:
Abstract

Objective: To compare 0.55 T and 3 T MRI for imaging patients with titanium total hip arthroplasty (THA). Patients with orthopedic metallic implants often require diagnostic imaging to evaluate adjacent tissues. MRI performance measures, including artifact levels and SNR, vary with field strength.

Methods: Six patients with titanium THA were scanned with similar protocols at 0.55 T and 3 T, including proton density (PD) weighted turbo spin echo (TSE), PD TSE with view-angle tilting (TSE + VAT), PD slice encoding for metal artifact correction (SEMAC), and short tau inversion recovery with SEMAC (STIR-SEMAC). Images from both field strengths were scored by two readers and qualitatively and quantitatively compared.

Results: Diagnostic confidence was significantly higher at 0.55 T compared to 3 T. Perceived metal artifact was substantially reduced at 0.55 T compared to 3 T. At 0.55 T, diagnostic imaging was achieved both without and with multi spectral imaging (MSI) for PD weighted images.

Conclusions: Compared to 3 T, 0.55 T MRI offers substantially reduced metal artifacts and higher diagnostic confidence when imaging titanium THA. Advanced multi-spectral techniques may not be required when the metallic components are entirely titanium.

Authors
Kübra Keskin, Sophia Cui, Bochao Li, Jordan Gross, Jay Acharya, Zorica Buser, Jay Lieberman, Brian Hargreaves, Krishna Nayak
Relevant Conditions

Hip Replacement