Continuous monitoring of irrigating fluid absorption during transurethral surgery.

Journal: Anaesthesia
Published:
Abstract

We have designed an automatic detector system for noninvasive monitoring of irrigating fluid absorption during transurethral resection of the prostate. Ethanol, which is used as a tracer for the fluid, is measured in the expired breath of the patient. The breath ethanol monitor can be used both in the awake patient and during general anaesthesia. A lap-top computer controls the monitor and calculates the fluid absorption 1-3 times per min and uses the pattern of ethanol changes to indicate whether the fluid is being taken by the intra- and/or extravascular route. The monitor has been used in routine operations for one year. The interpretation of absorption routes was checked by control methods or by evaluation of the postoperative breath-ethanol curve. We found the indicated absorption to be correct in all resections with absorption large enough for the monitor to present an interpretation.

Authors
R Hahn, H Larsson, T Ribbe
Relevant Conditions

Prostatectomy