Selective enrichment of 17 pyrethroids from lyophilised agricultural samples.

Journal: Journal Of Chromatography. A
Published:
Abstract

The screening of agricultural samples to determine 17 synthetic pyrethroids was investigated. Samples were lyophilised without losses of the insecticides, and then extracted with n-hexane. A simple, continuous preconcentration-elution system was developed, which included a silica sorbent column (packed with 50 mg) and used an air stream to carry the eluent (ethyl acetate) which minimised the eluate volume thus increasing the preconcentration factor; so no evaporation step was required. Pyrethroids were determined by gas chromatography-electron capture detection (GC-ECD) by using a 5% phenylmethylpolysiloxane-coated fused-silica capillary column; gas chromatography-mass spectrometry was used to identify the pyrethroids detected by GC-ECD monitoring. Limits of detection varied between 0.1 and 0.8 ng/ml (except for piperonyl butoxide, 25 ng/ml) with linear ranges from 1 to 200 ng/ml; the precision of the method was high (3-6%). Recoveries of 17 insecticides from 14 different agricultural samples fortified at levels of 20-100 ng/g ranged from 66 to 102% (bifenthrin and deltamethrin were those providing the lowest values, 66-87%). Pyrethroids were detected in eight samples (from the 100 unfortified agricultural samples tested) at concentrations lower than the established maximum residue limits (MRLs).

Authors
A Columé, S Cárdenas, M Gallego, M Valcárcel