Effect of steam sterilization inside the turbine chambers of dental turbines.
Objective: It has been demonstrated that contamination of the insides of high-speed dental turbines occurs and that bacteria as well as viruses may remain infectious when expelled from such turbines during subsequent use. Consequently, it has been widely recommended that a high-speed turbine be sterilized after each patient. The purpose of this study was to evaluate the effect of steam autoclaving on a high-speed dental turbine with a contaminated turbine chamber.
Methods: Streptococcus salivarius and endospores of Bacillus stearothermophilus were used as test organisms to determine the effectiveness of 4 different small non-vacuum autoclaves and one vacuum autoclave.
Results: The study demonstrated different efficiencies among the small non-vacuum autoclaves, the best showing close to a 6 log reduction of the test organisms inside the turbine chamber. When cleaning and lubrication of the high-speed dental turbine was carried out before autoclaving, this level of reduction was observed for all the examined non-vacuum autoclaves.
Conclusions: It is concluded that cleaning before sterilization is essential for safe use of high-speed dental turbines and that small non-vacuum autoclaves should be carefully evaluated before being used for the reprocessing of hollow instruments such as high-speed turbines.