Cobalt(II)-selective membrane electrode based on a recently synthesized benzo-substituted macrocyclic diamide.
Journal: Analytical Sciences : The International Journal Of The Japan Society For Analytical Chemistry
Published:
Abstract
A PVC-membrane electrode based on a recently synthesized 18-membered macrocyclic diamide is presented. The electrode reveals a Nernstian potentiometric response for Co2+ over a wide concentration range (2.0 x 10(-6)-1.0 x 10(-2) M). The electrode has a response time of about 10 s and can be used for at least 2 months without any divergence. The proposed sensor revealed very good selectivities for Co2+ over a wide variety of other metal ions, and could be used over a wide pH range (3.0-8.0). The detection limit of the sensor is 6.0 x 10(-7) M. It was successfully applied to the direct determination and potentiometric titration of cobalt ion.
Authors
M Shamsipur, T Poursaberi, S Rouhani, K Niknam, H Sharghi, M Ganjali