Long-term durability of the Hancock II porcine bioprosthesis.

Journal: The Journal Of Thoracic And Cardiovascular Surgery
Published:
Abstract

Background: Survival and prosthetic complications of patients receiving the Hancock II second-generation bioprosthesis (Medtronic, Inc, Minneapolis, Minn) in the aortic, mitral, mitral-aortic, and tricuspid positions were analyzed at 15 years' follow-up.

Methods: Between May 1983 and December 1993, 212 patients (104 men and 108 women; mean age, 63 +/- 9 years; age range, 29-81 years) received 66 aortic, 114 mitral, 26 mitral-aortic, and 6 tricuspid Hancock II valves. Thirty-one percent of patients had previous valve operations, 15% had concomitant cardiac procedures, and 87% were in New York Heart Association class III or IV. Follow-up included 1704 patient-years and was 98% complete, with a median of 9 patient-years (range, 0.013-17.4 years). Forty-six patients were at risk at 14 to 15 years, and 30 were at risk after 15 years.

Results: One hundred twenty-two (57%) of 212 patients died, 20 of them perioperatively. Fifteen-year actuarial Kaplan-Meier survival was 35.2% +/- 3.8%, and freedom from valve-related mortality was 84% +/- 3.5%, with no difference on the basis of position or age (<65 or >or=65 years). Percentages for freedom from thromboembolism, anticoagulant-related hemorrhage, endocarditis, and paravalvular leak were, respectively, 78.2% +/- 4%, 83.5% +/- 3.6%, 95.7% +/- 2%, and 97.3% +/- 1.4%, with no significant difference between the aortic and mitral positions. Freedom from structural valve deterioration was 71.8% +/- 5.6%: 88.9% +/- 6.2% in the aortic position versus 59.5% +/- 3.9% in the mitral position (P =.01) and 64.3% +/- 3% in the mitral-aortic position. In patients younger than 65 years, actual freedom from structural valve deterioration was less than that seen in older patients (84.5% +/- 3.5% vs 95% +/- 3.0%) and was better in the aortic versus the mitral position (92% +/- 4.5% vs 82% +/- 4.2%).

Conclusions: The Hancock II porcine valve showed excellent 15-year durability. We recommend its use in patients 65 years of age, as well as in younger patients undergoing aortic replacement.

Authors
Giulio Rizzoli, Tomaso Bottio, Gaetano Thiene, Giuseppe Toscano, Dino Casarotto