A modified parasternal wire technique for prevention and treatment of sternal dehiscence.

Journal: The Annals Of Thoracic Surgery
Published:
Abstract

Background: Sternal dehiscence with or without mediastinitis is a devastating complication of median sternotomy. Various techniques of sternotomy closure including 'figure of eight' wire sutures, nylon bands, and custom-made titanium-H plates have been described. We have devised and tested a new method of sternal closure to prevent sternal wound complications in patients at high risk of sternal dehiscence.

Methods: 1336 patients underwent sternotomy for various cardiac operations from January 1996 to January 2002. Patients were divided into two groups. Group I consisted of 560 patients who did not have any high risk factors for sternal dehiscence and received a standard six wire closure. Group II comprised of patients at high risk of sternal dehiscence and were divided randomly into subgroup II A (n = 390), which included patients who had conventional sternal closure. While in subgroup II B (n = 386) patients had a modified parasternal wire closure according to the finalized protocol.

Results: Sternal instability was noticed in 1/560 and none had sternal dehiscence in group I, but 16/390 patients had sternal instability and 3/390 had sternal dehiscence in subgroup II A, whereas only one patient in high risk subgroup II B developed sternal dehiscence with mediastinitis and required a pectoral flap advancement for sternal closure.

Conclusions: Use of modified parasternal wire closure in patients with a high risk of sternal dehiscence is a safe, effective, technically easily reproducible, as well as economical, method of preventing and treating sternal dehiscence.

Authors
Rajeev Sharma, Deepak Puri, Bishnu Panigrahi, Inderjeet Virdi
Relevant Conditions

Mediastinitis