Comparison of intraocular tonometry using three different non-invasive tonometers in children.

Journal: Graefe's Archive For Clinical And Experimental Ophthalmology = Albrecht Von Graefes Archiv Fur Klinische Und Experimentelle Ophthalmologie
Published:
Abstract

Background: In childhood glaucoma, the correct determination of intraocular pressure (IOP) is crucial in clinical decision-making. We therefore investigated how intraocular tonometry readings correlate with readings from commonly used tonometers.

Methods: IOP was measured unilaterally in 20 children suffering from congenital (n = 7) or secondary glaucoma (n = 13), 10 minutes after the induction of general anaesthesia. The children were aged from 1 month to 17 years (mean age 4.3 years, median age 1.3 years). Non-invasive applanation (Tono-Pen XL, Perkins tonometer) and indentation tonometry (Schiötz tonometer) were performed in random order prior to intraocular tonometry with a 26-gauge needle connected to a pressure sensor. Linear regression analysis and the coefficients of variance (CV) were used to compare the data obtained from the various tonometers.

Results: Compared with intraocular pressure, the CV was 10% for the Tono-Pen XL, 17% for the Schiötz, and 19% for the Perkins tonometer. The coefficient of determination (r(2)) was 0.74 for Tono-Pen XL, 0.60 for Schiötz and 0.78 for Perkins tonometry. The IOP values obtained with the Tono-Pen XL scattered homogeneously around the intraocular IOP, while the Perkins and Schiötz tonometers underestimated intraocular measured IOP.

Conclusions: Of the three tonometers evaluated in this series of paediatric glaucoma patients, the Tono-Pen XL most closely reflected true IOP.

Authors
Julia Lasseck, Thomas Jehle, Nicolas Feltgen, Wolf Lagrèze