Flicker perimetry (CFF) in glaucoma diagnosis
Journal: Klinika Oczna
Published:
Abstract
Objective: Evaluation of CFF perimetry results in patients with suspected glaucoma.
Methods: Visual field was measured on Octopus 301 perimeter with option "flicker perimetry". 49 patients (98 eyes) with suspected glaucoma were tested. Standard W/W perimetry revealed glaucomatous visual field defect in one eye, but not in the second eye. Patients with cataract and diabetes were not taken into consideration.
Results: Defects in CFF perimetry tests were found in 71.4% of cases, but in W/W perimetry test in 50.0% of cases.
Conclusions: Perimetry CFF is useful for early detection and follow up of glaucoma. Indices MD and LV may be used for those purposes.
Authors
Andrzej Dudziński, Izabela Zawojska, Renata Kinasz