An evaluation of the effects of eyeball structure on ocular pulse amplitude in healthy subjects.

Journal: International Ophthalmology
Published:
Abstract

To evaluate the effects of eyeball structure on ocular pulse amplitude (OPA) measured using dynamic contour tonometer (DCT). In 86 eyes of 43 healthy subjects, we measured OPA and intraocular pressure (IOP) with DCT (DCT-IOP), IOP with Goldmann applanation tonometry (GAT-IOP), central corneal thickness (CCT), corneal thickness 2 mm (2 mmCT) and 4 mm (4 mmCT) apart from the center, corneal volume within a 3.5-mm radius from the corneal center, corneal curvature, anterior chamber depth, anterior chamber volume, and axial length (AL). OPA had a significant positive correlation with GAT-IOP (Pearson's r = 0.412, p < 0.001), DCT-IOP (r = 0.350, p < 0.001), and 4 mmCT (r = 0.244, p = 0.0231), and had a significant negative correlation with AL (r = -0.268, p = 0.0122). In a multiple linear regression analysis, AL and GAT-IOP were significantly associated with OPA. OPA measured with DCT is significantly influenced by several factors, such as IOP, peripheral corneal thickness (4 mmCT), and AL.