DIRECT PEDICLE SCREW INSERTION PULLOUT STRENGTH.

Journal: Acta Ortopedica Brasileira
Published:
Abstract

Objective: Study the in vitro pullout strength of SpineGuard/Zavation Dynamic Surgical Guidance Z-Direct Screw (DSG Screw), a screw pedicle designed to be inserted using a direct insertion technique.

Methods: DSG Screws of 5.5 mm and 6.5 mm were introduced into polyurethane blocks with a density of 10 PCF (0,16 g/cm3). According to the experimental group, screws were inserted without pilot hole, with pilot without tapping, undertapping and line-to-line tapping. Screw pullout tests were performed using a universal test machine after screw insertion into polyurethane blocks.

Results: Screws inserted directly into the polyurethane blocks without pilot hole and tapping showed a statistically higher pullout strength. Insertion of the screw without tapping or with undertapping increases the pullout screw strength compared to line-to-line tapping.

Conclusions: DSG Screw showed the highest pullout strength after its insertion without pilot hole and tapping. Level of Evidence V, Expert Opinion.

Authors
Rômulo Pinheiro, Ariane Zamarioli, Thibault Chandanson, Keri George, Antonio Shimano, Helton Luiz Defino
Relevant Conditions

Spinal Fusion